Data Foundry

Medical device recalls 2023

CRTD DTBA2D1G VIVA XT IS1 DF1 OUS GOLD — Class I medical device recall Z-1736-2023

Ongoing recall by Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF), reported 2023-06-28, distributed nationwide. There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J)

Recall numberZ-1736-2023
Product typeMedical device
ClassificationClass I
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmMedtronic Inc., Cardiac Rhythm and Heart Failure (CRHF), Mounds View, MN, United States
Recall initiated2023-05-10
Classified2023-06-16
Reported by FDA2023-06-28
DistributedNationwide (US)
Quantity8 units
UPC / GTIN / UDI-DI00763000071431
Serial numbersBLJ622075S, BLJ622076S, BLJ622585S, BLJ622586S, BLJ622587S, BLJ623836S, BLJ623837S, BLJ623838S
Model numbersDTBA2D1G

Product

CRTD DTBA2D1G VIVA XT IS1 DF1 OUS GOLD, Model Number DTBA2D1G; Implantable Cardioverter Defibrillators

Reason for recall

There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1736-2023. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.