CRT-D DTBB1QQ VIVA QUAD S IS4/DF4 US — Class I medical device recall Z-1744-2023
Ongoing recall by Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF), reported 2023-06-28, distributed nationwide. There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J)
| Recall number | Z-1744-2023 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF), Mounds View, MN, United States |
| Recall initiated | 2023-05-10 |
| Classified | 2023-06-16 |
| Reported by FDA | 2023-06-28 |
| Distributed | Nationwide (US) |
| Quantity | 193 units |
| UPC / GTIN / UDI-DI | 00643169720145 |
| Serial numbers | BLK205166H, BLK205258H, BLK205259H, BLK205260H, BLK205262H, BLK205263H, BLK205267H, BLK205268H, BLK205269H, BLK205272H, BLK205273H, BLK205275H, BLK205276H, BLK205279H, BLK205284H, BLK205285H, BLK205286H, BLK205287H, BLK205288H, BLK205289H, BLK205291H, BLK205292H, BLK205293H, BLK205295H, BLK205296H and 168 more |
| Model numbers | DTBB1QQ |
Product
CRT-D DTBB1QQ VIVA QUAD S IS4/DF4 US, Model Number DTBB1QQ; Implantable Cardioverter Defibrillators
Reason for recall
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1744-2023. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.