CRT-D DTBC2D1 BRAVA IS1/DF1 INTL — Class I medical device recall Z-1749-2023
Ongoing recall by Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF), reported 2023-06-28, distributed nationwide. There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J)
| Recall number | Z-1749-2023 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF), Mounds View, MN, United States |
| Recall initiated | 2023-05-10 |
| Classified | 2023-06-16 |
| Reported by FDA | 2023-06-28 |
| Distributed | Nationwide (US) |
| Quantity | 7,921 units |
| UPC / GTIN / UDI-DI | 00643169720367, 00643169980129 |
| Serial numbers | BLW606597S, BLW606598S, BLW606599S, BLW606929S, BLW606930S, BLW606932S, BLW606935S, BLW606937S, BLW606938S, BLW606939S, BLW606940S, BLW606941S, BLW606942S, BLW606943S, BLW606944S, BLW606945S, BLW606946S, BLW606947S, BLW606948S, BLW606949S, BLW606950S, BLW606952S, BLW606953S, BLW606954S, BLW606955S and 475 more |
| Model numbers | DTBC2D1 |
Product
CRT-D DTBC2D1 BRAVA IS1/DF1 INTL, Model Number DTBC2D1; Implantable Cardioverter Defibrillators
Reason for recall
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.