Data Foundry

Medical device recalls 2013

LifeShield — Class II medical device recall Z-1751-2013

Terminated recall by Hospira Inc., reported 2013-07-24, distributed nationwide. Hospira has received reports of customers experiencing air-in-line alarms related to administration sets that

Recall numberZ-1751-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmHospira Inc., Lake Forest, IL, United States
Recall initiated2013-06-18
Classified2013-07-18
Reported by FDA2013-07-24
Terminated2016-11-07
DistributedNationwide (US)
Countries outside the USCR
Quantity14,400 units
Reason classespackaging
Lot numbers87-188-5H

Product

LifeShield; Sterile; Rx only; List No. 19666-28; SYMBIQ PUMP SET, CONVERTIBLE PIN, PIGGYBACK, 2 CLAVES PORTS AND OPTION LOK, NON-DEHP; NOMINAL LENGTH: 97 IN (249 CM); PRIMING VOLUME: 14 ML; For the administration of I.V. fluids; Made in Costa Rica; Hospira, Inc., Lake Forest, IL 60045 USA The Hospira Symbiq Infusion System is a general infusion system which provides IV infusion across the general spectrum of clinical care. The Symbiq pumps are standard sized pumps which are typically facility-based infusion pumps which may be used to deliver continuous and/or intermittent infusions to a variety of patient populations.

Reason for recall

Hospira has received reports of customers experiencing air-in-line alarms related to administration sets that have a drip chamber with a ball valve. Preliminary investigation findings found that there is a potential for the ball valve to not seal the drip chamber as intended, leading to air in the line of the administration set.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1751-2013. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.