LifeShield — Class II medical device recall Z-1751-2013
Terminated recall by Hospira Inc., reported 2013-07-24, distributed nationwide. Hospira has received reports of customers experiencing air-in-line alarms related to administration sets that
| Recall number | Z-1751-2013 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Hospira Inc., Lake Forest, IL, United States |
| Recall initiated | 2013-06-18 |
| Classified | 2013-07-18 |
| Reported by FDA | 2013-07-24 |
| Terminated | 2016-11-07 |
| Distributed | Nationwide (US) |
| Countries outside the US | CR |
| Quantity | 14,400 units |
| Reason classes | packaging |
| Lot numbers | 87-188-5H |
Product
LifeShield; Sterile; Rx only; List No. 19666-28; SYMBIQ PUMP SET, CONVERTIBLE PIN, PIGGYBACK, 2 CLAVES PORTS AND OPTION LOK, NON-DEHP; NOMINAL LENGTH: 97 IN (249 CM); PRIMING VOLUME: 14 ML; For the administration of I.V. fluids; Made in Costa Rica; Hospira, Inc., Lake Forest, IL 60045 USA The Hospira Symbiq Infusion System is a general infusion system which provides IV infusion across the general spectrum of clinical care. The Symbiq pumps are standard sized pumps which are typically facility-based infusion pumps which may be used to deliver continuous and/or intermittent infusions to a variety of patient populations.
Reason for recall
Hospira has received reports of customers experiencing air-in-line alarms related to administration sets that have a drip chamber with a ball valve. Preliminary investigation findings found that there is a potential for the ball valve to not seal the drip chamber as intended, leading to air in the line of the administration set.