Data Foundry

Medical device recalls 2020

Synchron Systems Opiate 2000 ng Reagent — Class II medical device recall Z-1751-2020

Terminated recall by Beckman Coulter Inc., reported 2020-04-29, distributed nationwide. Cross-reactivity testing may not meet the claims indicated in the Opiate 2000 ng (OP2) Instructions For Use. T

Recall numberZ-1751-2020
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmBeckman Coulter Inc., Brea, CA, United States
Recall initiated2019-09-04
Classified2020-04-20
Reported by FDA2020-04-29
Terminated2021-04-28
DistributedNationwide (US)
Quantity1,610 units
UPC / GTIN / UDI-DI15099590222031
Lot numbersM904147, OP2
Model numbers475006

Product

Synchron Systems Opiate 2000 ng Reagent (OP2 1 x 250); catalogue no. 475006 for use in conjunction with UniCel DxC 600/800 System(s) and SYNCHRON Systems Drugs of Abuse Testing (DAT) Urine Calibrators. The assay provides a rapid screening procedure for presence of the analyte in urine.

Reason for recall

Cross-reactivity testing may not meet the claims indicated in the Opiate 2000 ng (OP2) Instructions For Use. The presence of certain compounds in urine may result in false positive or false negative patient test results.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1751-2020. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.