Connecting Tube — Class II medical device recall Z-1756-2018
Terminated recall by Cook Inc., reported 2018-05-16, distributed in AL, CA, GA, NE, NV, NY, TX. Cap-adapter assembly of the device may potentially not be adequately tightened, possibly leading to fluid leak
| Recall number | Z-1756-2018 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Cook Inc., Bloomington, IN, United States |
| Recall initiated | 2018-03-29 |
| Classified | 2018-05-08 |
| Reported by FDA | 2018-05-16 |
| Terminated | 2019-04-09 |
| Distributed | AL, CA, GA, NE, NV, NY, TX |
| Countries outside the US | JP, KR, TW |
| Quantity | 6,345 |
| Reason classes | packaging |
| Lot numbers | 7797164, 7844038, 7844039, 7844040, 7844041, 7844042, 7846434, 7846435, 7853869, 7853870, 7853871, 7853872, 7853873, 7853874, 7857491, 7857492, 7857493, 7857494, 7857495, 7857496, 7857497, 7857498, 7857499, 7857500, 7863742, 7863744, 7863745, 7863747, 7863748, 7863749, 7863750, 7863751, 7871644, 7871645, 7871646, 7871647, 7881937, 7881938, 7881939, 7881940 and 28 more |
Product
Connecting Tube - Female to Male Luer Lock, HPCT8.8-120-M-FM Product Usage: The connecting tube is intended to be a conduit between a fluid reservoir and a vascular catheter in the patient. The connecting tube may be used during angiographic procedures to connect radiographic contrast power injector to a vascular catheter, arterial pressure monitoring, and fluid infusion.
Reason for recall
Cap-adapter assembly of the device may potentially not be adequately tightened, possibly leading to fluid leakage from or air aspiration into the connecting tube.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1756-2018. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.