CRTD DTMA2Q1 CLARIA MRI QUAD OUS DF1 — Class I medical device recall Z-1759-2023
Ongoing recall by Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF), reported 2023-06-28, distributed nationwide. There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J)
| Recall number | Z-1759-2023 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF), Mounds View, MN, United States |
| Recall initiated | 2023-05-10 |
| Classified | 2023-06-16 |
| Reported by FDA | 2023-06-28 |
| Distributed | Nationwide (US) |
| Quantity | 1,658 units |
| UPC / GTIN / UDI-DI | 00643169837614, 00763000206239 |
| Serial numbers | RPN600478S, RPN600480S, RPN600595S, RPN600596S, RPN600597S, RPN600598S, RPN600601S, RPN600603S, RPN600605S, RPN600609S, RPN600610S, RPN600616S, RPN600617S, RPN600618S, RPN600620S, RPN600626S, RPN600627S, RPN600630S, RPN600631S, RPN600634S, RPN600636S, RPN600637S, RPN600638S, RPN600639S, RPN600640S and 475 more |
| Model numbers | DTMA2Q1 |
Product
CRTD DTMA2Q1 CLARIA MRI QUAD OUS DF1, Model Number DTMA2Q1; Implantable Cardioverter Defibrillators
Reason for recall
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.