Data Foundry

Medical device recalls 2024

Clinical Chemistry Calibration Serum — Class II medical device recall Z-1759-2024

Ongoing recall by Randox Laboratories Ltd., reported 2024-05-15, distributed nationwide. Randox Laboratories has identified that Inorganic Phosphate in Calibration Serum Level 3, CAL2351 & CAL103

Recall numberZ-1759-2024
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmRandox Laboratories Ltd., Crumlin (North), N/A, Ireland
Recall initiated2024-03-05
Classified2024-05-03
Reported by FDA2024-05-15
DistributedNationwide (US)
UPC / GTIN / UDI-DI05055273200966
Lot numbers1260UE, 1262UE, 1295UE, 1297UE, 1298UE, 1315UE, 1325UE, 1326UE
Model numbersCAL10388, CAL2351

Product

Clinical Chemistry Calibration Serum (Catalog number Cal2351) that is 20x5ml lyophilized human based serum covering 38 commonly used clinical chemistry tests.

Reason for recall

Randox Laboratories has identified that Inorganic Phosphate in Calibration Serum Level 3, CAL2351 & CAL10388, is running with a negative bias on RX Series instruments compared to other methods.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1759-2024. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.