CRTD DTMB1D1 AMPLIA MRI US DF1 — Class I medical device recall Z-1761-2023
Ongoing recall by Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF), reported 2023-06-28, distributed nationwide. There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J)
| Recall number | Z-1761-2023 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF), Mounds View, MN, United States |
| Recall initiated | 2023-05-10 |
| Classified | 2023-06-16 |
| Reported by FDA | 2023-06-28 |
| Distributed | Nationwide (US) |
| Quantity | 3,005 units |
| UPC / GTIN / UDI-DI | 00643169837690, 00763000059606 |
| Serial numbers | RPV203080H, RPV203081H, RPV203082H, RPV203720H, RPV203721H, RPV203726H, RPV203727H, RPV203728H, RPV203731H, RPV203732H, RPV203733H, RPV203734H, RPV203735H, RPV203736H, RPV203737H, RPV203738H, RPV203739H, RPV203741H, RPV203743H, RPV203748H, RPV203751H, RPV203754H, RPV203755H, RPV203756H, RPV203757H and 475 more |
| Model numbers | DTMB1D1 |
Product
CRTD DTMB1D1 AMPLIA MRI US DF1, Model Number DTMB1D1; Implantable Cardioverter Defibrillators
Reason for recall
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.