Alaris Pump — Class I medical device recall Z-1768-2019
Terminated recall by CareFusion 303, Inc., reported 2019-07-24, distributed nationwide. The bezel manufacturing process for the FR-110 plastic may have resulted in its weakening. A bezel with weaken
| Recall number | Z-1768-2019 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | CareFusion 303, Inc., San Diego, CA, United States |
| Recall initiated | 2018-04-13 |
| Classified | 2019-07-16 |
| Reported by FDA | 2019-07-24 |
| Terminated | 2023-12-26 |
| Distributed | Nationwide (US) |
| Quantity | 607,618 |
| Reason classes | packaging |
| Model numbers | 8100 |
Product
Alaris Pump, Model 8100, with bezel mechanical assemblies manufactured with FR-110 plastic resin. The following products are affected by this recall as they are pumps or pump assemblies with bezels manufactured between April 2011 and June 2017 with the FR-110 plastic. Alaris Pump Module serviced with LVP Mechanism Sub Assembly (P/N 10942012, P/N 49000007, and P/N 49000203), and Alaris Pump module Bezel Kit Assembly (P/N 10964559 and P/N 49000204). Product Usage: The Alaris Pump module is a large volume infusion pump offered under the Alaris System. The Alaris Pump module will deliver medication and fluids using the IV administration sets for continuous or intermittent delivery through clinically acceptable routes of administration such as intravenous or epidural. There are 6 bezel posts on the Alaris Pump bezel assembly.
Reason for recall
The bezel manufacturing process for the FR-110 plastic may have resulted in its weakening. A bezel with weakened plastic may, over time, lead to separation of the bezel post (recall issue) as well as other damage to the bezel (i.e. external cracking). The separation of one or more bezel posts may result in free flow, over infusion, under infusion or interruption of infusion.