Data Foundry

Medical device recalls 2019

Alaris Pump — Class I medical device recall Z-1768-2019

Terminated recall by CareFusion 303, Inc., reported 2019-07-24, distributed nationwide. The bezel manufacturing process for the FR-110 plastic may have resulted in its weakening. A bezel with weaken

Recall numberZ-1768-2019
Product typeMedical device
ClassificationClass I
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmCareFusion 303, Inc., San Diego, CA, United States
Recall initiated2018-04-13
Classified2019-07-16
Reported by FDA2019-07-24
Terminated2023-12-26
DistributedNationwide (US)
Quantity607,618
Reason classespackaging
Model numbers8100

Product

Alaris Pump, Model 8100, with bezel mechanical assemblies manufactured with FR-110 plastic resin. The following products are affected by this recall as they are pumps or pump assemblies with bezels manufactured between April 2011 and June 2017 with the FR-110 plastic. Alaris Pump Module serviced with LVP Mechanism Sub Assembly (P/N 10942012, P/N 49000007, and P/N 49000203), and Alaris Pump module Bezel Kit Assembly (P/N 10964559 and P/N 49000204). Product Usage: The Alaris Pump module is a large volume infusion pump offered under the Alaris System. The Alaris Pump module will deliver medication and fluids using the IV administration sets for continuous or intermittent delivery through clinically acceptable routes of administration such as intravenous or epidural. There are 6 bezel posts on the Alaris Pump bezel assembly.

Reason for recall

The bezel manufacturing process for the FR-110 plastic may have resulted in its weakening. A bezel with weakened plastic may, over time, lead to separation of the bezel post (recall issue) as well as other damage to the bezel (i.e. external cracking). The separation of one or more bezel posts may result in free flow, over infusion, under infusion or interruption of infusion.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1768-2019. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.