Data Foundry

Medical device recalls 2013

LIGACLIP¿ Endoscopic Rotating Multiple Clip Applier Convenience Kit — Class II medical device recall Z-1776-2013

Terminated recall by Ethicon Endo-Surgery Inc, reported 2013-07-31, distributed nationwide. Ethicon Endo-Surgery is initiating a voluntary recall for LIGACLIP¿ 10mm M/L Endoscopic Rotating Multiple Clip

Recall numberZ-1776-2013
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmEthicon Endo-Surgery Inc, Cincinnati, OH, United States
Recall initiated2013-04-12
Classified2013-07-19
Reported by FDA2013-07-31
Terminated2014-05-01
DistributedNationwide (US)
Countries outside the USAE, AR, BE, BH, BR, CA, CL, CN, CO, CR, CY, EC, EG, GR, HN, ID, IL, IN, JM, JO, JP, KR, KW, LB, MX, NI, NL, PA, PE, PH, PK, PY, QA, SA, SG, SI, SV, TH, TR, TT, TW, UY, VE, ZA
Quantity49,357
Reason classesdevice malfunction

Product

LIGACLIP¿ Endoscopic Rotating Multiple Clip Applier Convenience Kit (Flex Trays), 10mm M/L Product Usage: The LIGACLIP Endoscopic Rotating Multiple Clip Applier is a sterile, single patient use instrument designed to provide a means of ligation through ENDOPATH¿ surgical trocars. The instrument delivers 20 titanium clips that individually advance after each firing. The shaft is made of a low glare material that minimizes reflective distortion. It is designed to rotate 360¿ in either direction. The rotating knob is located to allow for a one-handed technique.

Reason for recall

Ethicon Endo-Surgery is initiating a voluntary recall for LIGACLIP¿ 10mm M/L Endoscopic Rotating Multiple Clip Applier (ER320) due to potential clip formation and feeding issues which may result in improper clip formation, vessel damage and/or insufficient occlusion of the vessel or other structure

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1776-2013. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.