BD Pyxis MedBank Cabinet Software v2 — Class II medical device recall Z-1785-2022
Ongoing recall by CareFusion 303, Inc., reported 2022-10-05, distributed nationwide. The automated dispensing cabinet software update included a .dll file that is not compatible with MedBank Cabi
| Recall number | Z-1785-2022 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | CareFusion 303, Inc., San Diego, CA, United States |
| Recall initiated | 2022-05-24 |
| Classified | 2022-09-23 |
| Reported by FDA | 2022-10-05 |
| Distributed | Nationwide (US) |
| Quantity | 686 cabinets |
| Reason classes | software |
Product
BD Pyxis MedBank Cabinet Software v2.1.2.9 Legacy MedBank Cabinets with Matrix Drawers include: . CUBEX MedFlex operating with Windows 7 . MedBank 200 Main operating with Windows 7 . MedBank 500 Main operating with Windows 7. An automated dispensing cabinet intended to securely store and dispense medications to health care providers in a clinical setting.
Reason for recall
The automated dispensing cabinet software update included a .dll file that is not compatible with MedBank Cabinet stations with MedCart Matrix drawers. The cabinet system will stop functioning (Matrix drawers will stop opening in the MedBank Application) leading to delay in access to medications or supplies needed in residential or procedural care facilities.