Multi-parameter Command Module — Class II medical device recall Z-1793-2026
Ongoing recall by Spacelabs Healthcare, Inc., reported 2026-04-15, distributed nationwide. Multi-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measuremen
| Recall number | Z-1793-2026 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Spacelabs Healthcare, Inc., Snoqualmie, WA, United States |
| Recall initiated | 2026-02-27 |
| Classified | 2026-04-07 |
| Reported by FDA | 2026-04-15 |
| Distributed | Nationwide (US) |
| Countries outside the US | AR, CA, FR, KW, MX, PA, PK, PL, SA, TW |
| Quantity | 1,790 |
| UPC / GTIN / UDI-DI | 10841522106415, 10841522106422 |
| Serial numbers | 10841522106415, 10841522106422, 1496, 1496-005908, 1496-010295, 1496-014881, 1496-014884, 1496-014895, 1496-020263, 1496-020462, 1496-034574, 1496-101134, 1496-103164, 1496-116983, 1496-123122, 1496-139989, 1496-140522, 1496-142064, 1496-152103, 1496-152116, 1496-152331, 1496-152369, 1496-159546, 1496-159552, 1496-162049 and 475 more |
| Model numbers | 91496 |
Product
Multi-parameter Command Module, Model 91496, optioned with Masimo or Nellcor
Reason for recall
Multi-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early, resulting in Bad Curve error with no CO measurement or Irregular Curve error with CO measurement not automatically excluded, which may result in inaccurate readings, or the need of additional readings, which may lead to delayed care and/or fluid overload.