VITROS — Class II medical device recall Z-1805-2014

Terminated recall by Ortho-Clinical Diagnostics, reported 2014-06-25, distributed nationwide. Calibrator values assigned for the calibrator kit lots were released using a reference method that was undergo

Recall numberZ-1805-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmOrtho-Clinical Diagnostics, Rochester, NY, United States
Recall initiated2014-05-12
Classified2014-06-16
Reported by FDA2014-06-25
Terminated2017-03-23
DistributedNationwide (US)
Countries outside the USAU, BR, CA, CL, CN, CO, DE, ES, FR, GB, IN, IT, JP, MX, PA, SG, VE
Lot numbers91131
Model numbers6801873
Expiration dates2015-03-07, 2016-02-28

Product

VITROS (R) Chemistry Products FS Calibrator 1, Product Code/ Catalog Number 6801873 -- For in vitro diagnostic use only --- VITROS Chemistry Products FS Calibrator 1 is used in conjunction with VITROS Chemistry Products Calibrator Kits to calibrate the VITROS 5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.

Reason for recall

Calibrator values assigned for the calibrator kit lots were released using a reference method that was undergoing validation for a new lot of European Reference Material (ERM). The validation process was not complete prior to calibrator value assignment for VITROS (R) Chemistry Products hsCRP Reagent (Product Code 6801739), Generation (GEN) 28.