ICD-VR DVFC3D1 VISIA AF MRI S OUS/US DF1 — Class I medical device recall Z-1807-2023
Ongoing recall by Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF), reported 2023-06-28, distributed nationwide. There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J)
| Recall number | Z-1807-2023 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF), Mounds View, MN, United States |
| Recall initiated | 2023-05-10 |
| Classified | 2023-06-16 |
| Reported by FDA | 2023-06-28 |
| Distributed | Nationwide (US) |
| Quantity | 5,622 units |
| UPC / GTIN / UDI-DI | 00643169837812, 00643169837829, 00763000060107, 00763000206369 |
| Serial numbers | CWJ200335H, CWJ200337H, CWJ200338H, CWJ200390H, CWJ200391H, CWJ200392H, CWJ200394H, CWJ200395H, CWJ200398H, CWJ200400H, CWJ200401H, CWJ200402H, CWJ200404H, CWJ200405H, CWJ200406H, CWJ200407H, CWJ200408H, CWJ200409H, CWJ200410H, CWJ200411H, CWJ200413H, CWJ200414H, CWJ200415H, CWJ200418H, CWJ200421H and 475 more |
| Model numbers | DVFC3D1 |
Product
ICD-VR DVFC3D1 VISIA AF MRI S OUS/US DF1, Model Number DVFC3D1; Implantable Cardioverter Defibrillators
Reason for recall
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.