ICD-VR DVMB2D1 EVERA MRI XT OUS DF1 — Class I medical device recall Z-1810-2023
Ongoing recall by Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF), reported 2023-06-28, distributed nationwide. There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J)
| Recall number | Z-1810-2023 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF), Mounds View, MN, United States |
| Recall initiated | 2023-05-10 |
| Classified | 2023-06-16 |
| Reported by FDA | 2023-06-28 |
| Distributed | Nationwide (US) |
| Quantity | 2,310 units |
| UPC / GTIN / UDI-DI | 00643169837768, 00763000206178, 00763000612030 |
| Serial numbers | CWE600295S, CWE600337S, CWE600338S, CWE600339S, CWE600340S, CWE600341S, CWE600342S, CWE600348S, CWE600349S, CWE600350S, CWE600351S, CWE600352S, CWE600353S, CWE600354S, CWE600355S, CWE600356S, CWE600357S, CWE600358S, CWE600359S, CWE600360S, CWE600361S, CWE600362S, CWE600363S, CWE600364S, CWE600365S and 475 more |
| Model numbers | DVMB2D1 |
Product
ICD-VR DVMB2D1 EVERA MRI XT OUS DF1, Model Number DVMB2D1; Implantable Cardioverter Defibrillators
Reason for recall
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.