Data Foundry

Medical device recalls 2017

Radiation Therapy Treatment Planning System — Class II medical device recall Z-1815-2017

Terminated recall by RAYSEARCH LABORATORIES AB, reported 2017-04-19, distributed nationwide. An error in NVIDIA GPU (Graphics Processing Unit) card drivers can occur for certain software programs. In a n

Recall numberZ-1815-2017
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmRAYSEARCH LABORATORIES AB, Stockholm, N/A, Sweden
Recall initiated2017-01-18
Classified2017-04-12
Reported by FDA2017-04-19
Terminated2019-06-12
DistributedNationwide (US)
Quantity741 units
Reason classessoftware
Model numbers4.5, 4.7, 5.0

Product

Radiation Therapy Treatment Planning System, Model 4.5, 4.7, 5.0 RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs

Reason for recall

An error in NVIDIA GPU (Graphics Processing Unit) card drivers can occur for certain software programs. In a non-standard Citrix environment where the GPU settings are configured with a virtual Citrix graphics board (graphics card) an erroneous calculation may be allowed.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1815-2017. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.