TotalCare Bed Systems — Class II medical device recall Z-1816-2017

Terminated recall by Hill-Rom, Inc., reported 2017-04-19, distributed nationwide. The bed exit alarm may not function properly if the alarm has been armed for a period longer than seventy-two

Recall numberZ-1816-2017
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmHill-Rom, Inc., Batesville, IN, United States
Recall initiated2017-02-23
Classified2017-04-12
Reported by FDA2017-04-19
Terminated2017-09-19
DistributedNationwide (US)
Countries outside the USAE, AU, BD, BR, CA, CN, CY, DE, GR, IN, IT, KW, NL, PK, SA, TH
Quantity3,037
Model numbersP1900, P1947A, P1948A, PR1900, U1900

Product

TotalCare Bed Systems, Model Numbers P1900, PR1900, and U1900 with 3-Level Patient Position Monitoring and Scale Upgrade Kits, Model Numbers P1947A and P1948A).

Reason for recall

The bed exit alarm may not function properly if the alarm has been armed for a period longer than seventy-two (72) hours without the alarm being triggered by patient movement or removal of patient weight within that time period.