TotalCare Bed Systems — Class II medical device recall Z-1816-2017
Terminated recall by Hill-Rom, Inc., reported 2017-04-19, distributed nationwide. The bed exit alarm may not function properly if the alarm has been armed for a period longer than seventy-two
| Recall number | Z-1816-2017 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Hill-Rom, Inc., Batesville, IN, United States |
| Recall initiated | 2017-02-23 |
| Classified | 2017-04-12 |
| Reported by FDA | 2017-04-19 |
| Terminated | 2017-09-19 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AU, BD, BR, CA, CN, CY, DE, GR, IN, IT, KW, NL, PK, SA, TH |
| Quantity | 3,037 |
| Model numbers | P1900, P1947A, P1948A, PR1900, U1900 |
Product
TotalCare Bed Systems, Model Numbers P1900, PR1900, and U1900 with 3-Level Patient Position Monitoring and Scale Upgrade Kits, Model Numbers P1947A and P1948A).
Reason for recall
The bed exit alarm may not function properly if the alarm has been armed for a period longer than seventy-two (72) hours without the alarm being triggered by patient movement or removal of patient weight within that time period.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1816-2017. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.