Philips Digital Diagnost Software — Class II medical device recall Z-1818-2012
Terminated recall by Philips Healthcare Inc., reported 2012-06-27, distributed nationwide. With Digital Diagnost (DIDI) R 2.0.2 a mirrored "R in a circle" is burnt into the image (if mirrored) in the l
| Recall number | Z-1818-2012 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Philips Healthcare Inc., Andover, MA, United States |
| Recall initiated | 2012-05-22 |
| Classified | 2012-06-15 |
| Reported by FDA | 2012-06-27 |
| Terminated | 2017-02-24 |
| Distributed | Nationwide (US) |
| Countries outside the US | AE, AF, AR, AT, AU, AW, BB, BD, BE, BG, BH, BO, BR, BS, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GH, GR, GT, GY, HK, HN, HR, HU, ID, IE, IL, IN, IQ, IR, IS, IT, JM, JO, JP, KE, KR, KW, LB, LK, LT, LU, LV, MA, MT, MX, MY, NG, NI, NL, NO, NP, NZ, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RS, RU, SA, SE, SG, SI, SK, SV, TH, TN, TR, TT, TW, UA, UY, VE, VN, ZA |
| Quantity | 752 units |
Product
Philips Digital Diagnost Software, software release 2.0.2 and 2.0.2SP1 including mirror icon Stationary Fluoroscopic system
Reason for recall
With Digital Diagnost (DIDI) R 2.0.2 a mirrored "R in a circle" is burnt into the image (if mirrored) in the lower right corner of the image. When such an image is mirrored back inside the PACS system, an unmirrored "R in a circle" appears in the lower left corner. This can be mistaken for a "right patient side" marker, although this can appear on the left patient side
Machine access