Data Foundry

Medical device recalls 2024

VITROS Immunodiagnostic Products Free T3 Reagent Pack-IVD is for the… — Class II medical device recall Z-1818-2024

Ongoing recall by QUIDELORTHO, reported 2024-05-22, distributed nationwide. T3 Reagent Packs used with Free T3 Calibrators may produce positively biased results for patient and quality c

Recall numberZ-1818-2024
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmQUIDELORTHO, Rochester, NY, United States
Recall initiated2024-04-09
Classified2024-05-14
Reported by FDA2024-05-22
DistributedNationwide (US)
Countries outside the USAU, BE, BM, BR, CA, CL, CN, CO, DE, DK, ES, FR, GB, IN, IT, JP, MX, NL, NO, PT, RU, SE, SG
Quantity21,268 units
UPC / GTIN / UDI-DI10758750008797
Lot numbers3042, 3052, 3061, 3080, 3090, 3100, 3110, 3126
Expiration dates2024-04-09, 2024-05-02, 2024-06-06, 2024-06-14, 2024-07-29, 2024-08-13, 2024-09-04

Product

VITROS Immunodiagnostic Products Free T3 Reagent Pack-IVD is for the quantitative measurement of free triiodothyronine (FT3) in human serum and plasma (EDTA or heparin). Product Code: 131 5589

Reason for recall

T3 Reagent Packs used with Free T3 Calibrators may produce positively biased results for patient and quality control samples. The positive bias may cause erroneous patient results or a delay in testing that may impact patient management, clinical assessment, and the results of other thyroid function tests.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1818-2024. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.