VITROS Immunodiagnostic Products Free T3 Calibrators-IVD are for use… — Class II medical device recall Z-1819-2024

Ongoing recall by QUIDELORTHO, reported 2024-05-22, distributed nationwide. Free T3 Calibrators used with T3 Reagent Packs may produce positively biased results for patient and quality c

Recall numberZ-1819-2024
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmQUIDELORTHO, Rochester, NY, United States
Recall initiated2024-04-09
Classified2024-05-14
Reported by FDA2024-05-22
DistributedNationwide (US)
Countries outside the USAU, BE, BM, BR, CA, CL, CN, CO, DE, DK, ES, FR, GB, IN, IT, JP, MX, NL, NO, PT, RU, SE, SG
UPC / GTIN / UDI-DI10758750008438
Lot numbers3042, 3052, 3061, 3080, 3090, 3100, 3110, 3126
Expiration dates2024-04-09, 2024-05-02, 2024-06-06, 2024-06-14, 2024-07-29, 2024-08-13, 2024-09-04

Product

VITROS Immunodiagnostic Products Free T3 Calibrators-IVD are for use with Free T3 Reagent Pack for the quantitative measurement of free triiodothyronine (FT3) in human serum and plasma (EDTA or heparin). Product Code: 111 2820

Reason for recall

Free T3 Calibrators used with T3 Reagent Packs may produce positively biased results for patient and quality control samples. The positive bias may cause erroneous patient results or a delay in testing that may impact patient management, clinical assessment, and the results of other thyroid function tests.