Allura Xper Release 8 — Class II medical device recall Z-1820-2017
Terminated recall by Philips Electronics North America Corporation, reported 2017-04-26, distributed nationwide. Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If
| Recall number | Z-1820-2017 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Philips Electronics North America Corporation, Andover, MA, United States |
| Recall initiated | 2017-03-22 |
| Classified | 2017-04-14 |
| Reported by FDA | 2017-04-26 |
| Terminated | 2021-05-17 |
| Distributed | Nationwide (US) |
| Countries outside the US | CA |
| Quantity | 6,992 |
| Reason classes | packaging |
Product
Allura Xper Release 8.2 (Laird chiller is always installed); Allura Xper FD10 722026; Allura Xper FD10/10 722027; Allura Xper FD20/10 biplane 722029; Allura Xper FD10 OR Table 722033; Allura Xper FD10/10 OR Table 722034; Allura Xper FD20/10 OR Table 722036; Allure Xper FD20/20 722038; Allura Xper FD20/20 biplane OR Table 722039
Reason for recall
Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.