ICD COBALT VR MRI IS1 DF1 — Class I medical device recall Z-1820-2023
Ongoing recall by Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF), reported 2023-06-28, distributed nationwide. There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J)
| Recall number | Z-1820-2023 |
|---|---|
| Product type | Medical device |
| Classification | Class I |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF), Mounds View, MN, United States |
| Recall initiated | 2023-05-10 |
| Classified | 2023-06-16 |
| Reported by FDA | 2023-06-28 |
| Distributed | Nationwide (US) |
| Quantity | 1,930 units |
| UPC / GTIN / UDI-DI | 00763000178499, 00763000178505 |
| Serial numbers | RSE600007S, RSE600009S, RSE600010S, RSE600011S, RSE600012S, RSE600013S, RSE600014S, RSE600015S, RSE600016S, RSE600017S, RSE600019S, RSE600020S, RSE600023S, RSE600024S, RSE600025S, RSE600028S, RSE600030S, RSE600033S, RSE600034S, RSE600036S, RSE600041S, RSE600042S, RSE600044S, RSE600046S, RSE600048S and 475 more |
| Model numbers | DVPB3D1 |
Product
ICD COBALT VR MRI IS1 DF1, Model Number DVPB3D1; Implantable Cardioverter Defibrillators
Reason for recall
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.