Cardinal Steam Biological Indicator — Class II medical device recall Z-1823-2013
Terminated recall by Steritec Products Mfc. Co., Inc., reported 2013-08-07, distributed nationwide. Mesa Labs, based out of Bozeman, MT, informed SteriTec Products, Inc. of a product recall involving the EZTest
| Recall number | Z-1823-2013 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Steritec Products Mfc. Co., Inc., Englewood, CO, United States |
| Recall initiated | 2013-06-10 |
| Classified | 2013-07-29 |
| Reported by FDA | 2013-08-07 |
| Terminated | 2013-12-19 |
| Distributed | Nationwide (US) |
| Countries outside the US | AR, AU, CA, CZ, IT, TR |
| Quantity | 200 |
| Reason classes | labeling |
| Lot numbers | S-444 |
Product
Cardinal Steam Biological Indicator, SteriTec Product Code PL 218, Distributor Product Code T40360EZ EZTest Steam Biological Indicator: Mesa EZTest Steam Biological Indicator is a self contained Biological Indicator intended for monitoring the efficacy of 121¿C, 132¿C, 134¿C and 135¿C steam sterilization cycles. Mesa EZTest has a recommended incubation time of 24 hours.
Reason for recall
Mesa Labs, based out of Bozeman, MT, informed SteriTec Products, Inc. of a product recall involving the EZTest Steam and Smart-Read Biological Indicators (BI) because the readout time is not in specification with the label claims.