Medline Med-Checks BI Test packs with Instant Readout Integrator — Class II medical device recall Z-1828-2013
Terminated recall by Steritec Products Mfc. Co., Inc., reported 2013-08-07, distributed nationwide. Mesa Labs, based out of Bozeman, MT, informed SteriTec Products, Inc. of a product recall involving the EZTest
| Recall number | Z-1828-2013 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Steritec Products Mfc. Co., Inc., Englewood, CO, United States |
| Recall initiated | 2013-06-10 |
| Classified | 2013-07-29 |
| Reported by FDA | 2013-08-07 |
| Terminated | 2013-12-19 |
| Distributed | Nationwide (US) |
| Countries outside the US | AR, AU, CA, CZ, IT, TR |
| Quantity | 3,535 units |
| Reason classes | labeling |
| Lot numbers | 130401 |
Product
Medline Med-Checks BI Test packs with Instant Readout Integrator, SteriTec Product Code PL 200, Distributor Product Code MDS200500IR SteriTec Steam Biological Test Packs with Instant Readout Integrator Card: The Steam Biological Test Pack with Instant Readout Integrator is a steam sterilization monitor designed specifically for biological testing of pre-vacuum steam sterilizers operating at 132¿C (270¿F) for 3 minutes or longer. The Instant Readout Integrator Card gives the operator a prediction of the biological test outcome
Reason for recall
Mesa Labs, based out of Bozeman, MT, informed SteriTec Products, Inc. of a product recall involving the EZTest Steam and Smart-Read Biological Indicators (BI) because the readout time is not in specification with the label claims.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1828-2013. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.