Data Foundry

Medical device recalls 2016

CADD Administration Sets with Flow Stop — Class II medical device recall Z-1831-2016

Terminated recall by Smiths Medical ASD, Inc., reported 2016-06-08, distributed in DC, PR. Smiths Medical became aware that under delivery of medication may occur on CADD Administration Sets with Flow

Recall numberZ-1831-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSmiths Medical ASD, Inc., Saint Paul, MN, United States
Recall initiated2016-04-22
Classified2016-05-27
Reported by FDA2016-06-08
Terminated2017-12-13
DistributedDC, PR
Countries outside the USAE, AT, BE, CA, CN, DE, DK, ES, FI, FR, GB, HK, IE, IN, IT, JP, LU, NZ, PA, PT, SA, SE, SG, TW, ZA
Quantity13,696,830 units

Product

CADD Administration Sets with Flow Stop, Product Reorder No: 21-7321-01, 21-7321-24, 21-7322-01, 21-7322-24, 21-7323-24, 21-7324-01, 21-7324-24, 21-7333-24, 21-7336-01, 21-7336-24, 21-7339-01, 21-7339-24, 21-7359-01, 21-7359-24, 21-7383-01, 21-7383-24, 21-7390-01, 21-7390-24, 21-7391-01, 21-7391-24, 21-7394-01, 21-7394-24, 21-7395-24. The CADD Medication Cassette Reservoirs with Flow Stop are a modification to the current CADD Medication Cassette Reservoirs. The reservoirs will incorporate a set-based free flow protection component (i.e. Flow Stop) that is designed to occlude the tube if the reservoirs are accidentally placed onto the pump incorrectly or become detached from the pump. The Flow Stop will be located on the reservoir housing, which is attached to the pump. The reservoir will be provided to the user in an open state. Before the reservoir can be attached to the pump, the blue "CLIP" must be removed to activate the Flow Stop. However, after attaching the reservoir to the pump, the user can still remove the reservoir from the pump and prime it by holding the Flow Stop in the open position.

Reason for recall

Smiths Medical became aware that under delivery of medication may occur on CADD Administration Sets with Flow Stop.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1831-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.