Data Foundry

Medical device recalls 2015

Artis zee/ zeego systems — Class II medical device recall Z-1846-2015

Terminated recall by Siemens Medical Solutions USA, Inc, reported 2015-07-01, distributed nationwide. There exists a possible position sensor fault in the swivel base axis not being detected by the system softwar

Recall numberZ-1846-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSiemens Medical Solutions USA, Inc, Malvern, PA, United States
Recall initiated2015-05-22
Classified2015-06-25
Reported by FDA2015-07-01
Terminated2016-12-13
DistributedNationwide (US)
Quantity475
Reason classessoftware
Serial numbers124002, 124006, 124007, 124015, 124034, 124039, 130013, 131000, 131005, 131006, 131008, 131011, 131013, 135000, 135117, 135120, 135129, 135132, 135137, 135143, 135148, 135151, 135152, 135157, 135169 and 450 more
Model numbers10094135, 10094139, 10094141, 10094142, 10094143

Product

Artis zee/ zeego systems; Product Usage: Usage: Artis zee/ zeego systems are dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed include cardiac angiography, neuro angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. .

Reason for recall

There exists a possible position sensor fault in the swivel base axis not being detected by the system software. When moving, the system could potentially exceed its usual speed, resulting in a collision. No injuries reported.