Data Foundry

Medical device recalls 2014

SteriGear Urinary Drain Bag with Fig Leaf Cover — Class II medical device recall Z-1848-2014

Terminated recall by SteriGear, LLC, reported 2014-07-02, distributed nationwide. SteriGear notifed customers of circumstances in which The Fig Leaf urinary drain bag valve can close creating

Recall numberZ-1848-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSteriGear, LLC, Provo, UT, United States
Recall initiated2011-01-11
Classified2014-06-23
Reported by FDA2014-07-02
Terminated2015-01-02
DistributedNationwide (US)
Quantity110,560
UPC / GTIN / UDI-DI00851616002007
Model numbers10270

Product

SteriGear Urinary Drain Bag with Fig Leaf Cover, 2000 ml, REF 10270, The Fig Leaf, SteriGear 362 S. University Avenue, provo, UT 84601, 1-800-398-3259, SteriGear.com, Sterile, EO Urinary Drainage Collection Kit, for Indwelling Catheter. This device is intended to collect the Urine drained from an indwelling catheter

Reason for recall

SteriGear notifed customers of circumstances in which The Fig Leaf urinary drain bag valve can close creating a negative pressure that can suck the non-return valve closed.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1848-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.