Data Foundry

Medical device recalls 2014

Spacelabs Healthcare Qube Compact Monitor — Class II medical device recall Z-1852-2014

Terminated recall by Spacelabs Healthcare Inc, reported 2014-07-02, distributed in AL, AZ, CA, CO, FL, GA, IA, ID…. Spacelabs Healthcare is recalling Qube Compact Monitor, Model 91390, due to a potential chance of losing netwo

Recall numberZ-1852-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmSpacelabs Healthcare Inc, Snoqualmie, WA, United States
Recall initiated2014-06-06
Classified2014-06-23
Reported by FDA2014-07-02
Terminated2015-02-17
DistributedAL, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MS, MT, NC, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, WA, WI, WY
Model numbers91390

Product

Spacelabs Healthcare Qube Compact Monitor, Model 91390. Intended to alert the user to alarm conditions that are reported by Spacelabs Healthcare parameter modules and/or other physiologic monitors via Flexport interfaces.

Reason for recall

Spacelabs Healthcare is recalling Qube Compact Monitor, Model 91390, due to a potential chance of losing network connection after switching from hardwired Ethernet connection to using the Qube Docking Station without powering down the device. The bedside monitor and all of its alarms will continue to function normally. No one has been injured as a result of this issue.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1852-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.