Ultrasonic Cleaning Systems Triton 36 Model Number 6003000804 Triton… — Class II medical device recall Z-1859-2025
Ongoing recall by Ultra Clean Systems, Inc., reported 2025-06-04, distributed nationwide. During routine preventative maintenance checks, it was revealed that carbon filter sensors were breaking down
| Recall number | Z-1859-2025 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Ultra Clean Systems, Inc., Oldsmar, FL, United States |
| Recall initiated | 2025-03-31 |
| Classified | 2025-05-28 |
| Reported by FDA | 2025-06-04 |
| Distributed | Nationwide (US) |
| Countries outside the US | CA |
| Quantity | 96 systems |
| Reason classes | specification failure |
| UPC / GTIN / UDI-DI | 00816634020006, 00816634020013 |
| Serial numbers | 1511-XL72-230-1, 1601-XL36-249-1, 1604-XL36-277-1, 1606-XL72-291-1, 1706-XL72TR-401-1, 1805-XL72TR-544-1, 1805-XL72TR-546-1, 1806-XL72TR-564-1, 1806-XL72TR-565-1, 1807-XL72TR-568-1, 1808-XL36TR-596-1, 1809-XL72TR-611-1, 1810-XL36TR-618-1, 1905-XL36TR-707-1, 1907-XL36TR-721-1, 1910-XL72TR-788-1, 1911-XL36TR-800-1, 2001-XL36TR-834-1, 2002-XL72TR-844-1, 2004-XL36TR-859-1, 2004-XL36TR-873-1, 2004-XL72TR-878-1, 2004-XL72TR-879-1, 2005-XL72TR-881-1, 2006-XL72TR-890-1 and 71 more |
| Model numbers | 6003000804, 6003000805 |
Product
Ultrasonic Cleaning Systems Triton 36 Model Number 6003000804 Triton 72 Model Number 6003000805 Ultrasonic cleaning system
Reason for recall
During routine preventative maintenance checks, it was revealed that carbon filter sensors were breaking down due to electrochemical degradation.