Data Foundry

Medical device recalls 2014

LAP LASER PICTOR 3D Multi Color Laser System PICTOR 3D is a LAP laser… — Class II medical device recall Z-1866-2014

Terminated recall by LAP of America Laser Applications, LLC, reported 2014-07-02, distributed in NC. While projecting yellow lines it was notice that the yellow projection has the potential to splits into two se

Recall numberZ-1866-2014
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmLAP of America Laser Applications, LLC, Boynton Beach, FL, United States
Recall initiated2014-03-21
Classified2014-06-24
Reported by FDA2014-07-02
Terminated2016-09-26
DistributedNC
Quantity1 systems

Product

LAP LASER PICTOR 3D Multi Color Laser System PICTOR 3D is a LAP laser system which uses multi projection format that defines isocenters, field and MLC contours along with a 3D patient view.

Reason for recall

While projecting yellow lines it was notice that the yellow projection has the potential to splits into two separate red and green projections. In this case the difference between the red and green projection was up to 2.5mm.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1866-2014. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.