Data Foundry

Medical device recalls 2021

Electrostatic Filter Angled — Class II medical device recall Z-1866-2021

Ongoing recall by Covidien, LP, reported 2021-06-23, distributed nationwide. The company was notified by their supplier of sterilization services for airway products in Italy of potential

Recall numberZ-1866-2021
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmCovidien, LP, North Haven, CT, United States
Recall initiated2021-04-29
Classified2021-06-16
Reported by FDA2021-06-23
DistributedNationwide (US)
Quantity32,625
Reason classeschemical contamination
UPC / GTIN / UDI-DI30884522012984, 40884522012981
Lot numbers20G1304FAX, 20G1305FAX, 20L0026FAX, 20L0027FAX, 20L0410FAX, 20L0411FAX, 20L0412FAX
Model numbers350S19006

Product

Electrostatic Filter Angled, Item Code 350S19006 - Product Usage: indicated for single use on anesthetized patients and respiratory care patients who require a breathing circuit with ISO standard 15mm or 22 mm connectors.

Reason for recall

The company was notified by their supplier of sterilization services for airway products in Italy of potential deviations from validated parameters for ethylene oxide sterilization.