Data Foundry

Medical device recalls 2016

Novocastr Liquid Mouse Monoclonal Antibody Renal Cell Carcinoma Marker — Class II medical device recall Z-1876-2016

Terminated recall by Leica Microsystems, Inc., reported 2016-06-08, distributed in AR, CA, CT, FL, ID, IL, IN, MT…. Testing of retained units indicated that Lot numbers of CDX2 and RCC indicated are not stable up to the expiry

Recall numberZ-1876-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmLeica Microsystems, Inc., Buffalo Grove, IL, United States
Recall initiated2016-04-05
Classified2016-06-01
Reported by FDA2016-06-08
Terminated2017-11-30
DistributedAR, CA, CT, FL, ID, IL, IN, MT, OH, PA, TN, TX
Countries outside the USCA
Quantity270 units
Reason classeslabeling
Lot numbers6023543, 6032345, 6036120

Product

Novocastr Liquid Mouse Monoclonal Antibody Renal Cell Carcinoma Marker (RCC) is recommended for use as part of a panel of antibodies in the definition of tumors of proximal nephrogenic differentiation.

Reason for recall

Testing of retained units indicated that Lot numbers of CDX2 and RCC indicated are not stable up to the expiry date on the product labeling.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1876-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.