Data Foundry

Medical device recalls 2017

BBL Trypticase Soy Agar with 5% Sheep Blood — Class III medical device recall Z-1876-2017

Terminated recall by Becton Dickinson & Co., reported 2017-05-03, distributed nationwide. BD has confirmed that one lot of TSA II and Chocolate Agar I-Plates may fail to recover Haemophilus influenzae

Recall numberZ-1876-2017
Product typeMedical device
ClassificationClass III
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmBecton Dickinson & Co., Sparks, MD, United States
Recall initiated2015-03-27
Classified2017-04-25
Reported by FDA2017-05-03
Terminated2017-04-26
DistributedNationwide (US)
Quantity1,600 plates
Reason classesdevice malfunction
Serial numbers4365741
Model numbers221302

Product

BBL Trypticase Soy Agar with 5% Sheep Blood (TSA II) and Chocolate II Agar (GC II Agar with Hemoglobin and IsoVitaleX)  I Plate Product Usage: Testing - Trypticase Soy Agar with 5% Sheep Blood is used for the growth of fastidious organisms and for the visualization of hemolytic reactions. Chocolate II Agar is an enriched medium for the isolation and cultivation of Neisseria species.

Reason for recall

BD has confirmed that one lot of TSA II and Chocolate Agar I-Plates may fail to recover Haemophilus influenzae 10211 and Haemophilus parainfluenza 51505 on the Chocolate Agar side of the plate. All other QC organisms are correctly recovered.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1876-2017. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.