Replacement HEPA filter for Natus Sterile-Drier Model 43 and Model 44 — Class II medical device recall Z-1878-2016
Terminated recall by Natus Medical Incorporated, reported 2016-06-08, distributed in AZ, CA, DC, FL, IL, KS, MA, MD…. The replacement HEPA Filters have been manufactured with sealant that may not be properly cured. Improperly cu
| Recall number | Z-1878-2016 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Natus Medical Incorporated, Seattle, WA, United States |
| Recall initiated | 2016-04-22 |
| Classified | 2016-06-01 |
| Reported by FDA | 2016-06-08 |
| Terminated | 2016-12-16 |
| Distributed | AZ, CA, DC, FL, IL, KS, MA, MD, MN, NC, NH, NY, OH, OK, PA, VA, WA, WI |
| Countries outside the US | CA, SG |
| Quantity | 21 filters |
| Reason classes | sterility |
| Model numbers | 100105-01, 54343, 54344 |
Product
Replacement HEPA filter for Natus Sterile-Drier Model 43 and Model 44. Filter Catalogue Part Number: 54343, 54344, and 100105-01.
Reason for recall
The replacement HEPA Filters have been manufactured with sealant that may not be properly cured. Improperly cured sealant can emit a toxic gas when heated, and can drip down during use, contaminating the interior chamber of the Sterile Drier and its contents.