Abbott Accelerator a3600 Centrifuge Module tube loading algorithm — Class II medical device recall Z-1888-2020
Ongoing recall by Inpeco S.A., reported 2020-05-13, distributed in AL, CA, CT, DE, FL, IA, KY, LA…. The FlexLab Centrifuge Module loading algorithm may lead to unbalanced loads which could damage the centrifuge
| Recall number | Z-1888-2020 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | Inpeco S.A., Lugano, N/A, Switzerland |
| Recall initiated | 2020-04-07 |
| Classified | 2020-05-05 |
| Reported by FDA | 2020-05-13 |
| Distributed | AL, CA, CT, DE, FL, IA, KY, LA, MA, MD, MI, MN, MO, MT, NC, NE, NH, NJ, NM, NY, OH, PA, RI, TN, TX, VA, WA |
| Countries outside the US | MX |
| Quantity | 23 units |
| Reason classes | software |
| Serial numbers | ACP.357, ACP.388, ACP.396, ACP.397, ACP.399, ACP.400, ACP.402, ACP.404, ACP.405, ACP.407, ACP.408, ACP.409, ACP.410, ACP.412, ACP.413, ACP.416, ACP.417, ACP.418, ACP.419, ACP.420, ACP.421, ACP.422, ACP.423 |
Product
Abbott Accelerator a3600 Centrifuge Module tube loading algorithm, Inpeco SA P/N: FLX-202, with Firmware Versions: CM_1-7-0.H86 or xCMd_2-5-0.elf and CM_1-7-1.H86 or xCMd_2-5-1-849fad9.elf
Reason for recall
The FlexLab Centrifuge Module loading algorithm may lead to unbalanced loads which could damage the centrifuge.