Data Foundry

Medical device recalls 2021

OSSUR MIAMI J SELECT COLLAR — Class II medical device recall Z-1896-2021

Terminated recall by Ossur Americas, reported 2021-06-23, distributed nationwide. Due to insufficient column strength of the racks for all load conditions, racks associated with height adjustm

Recall numberZ-1896-2021
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmOssur Americas, Foothill Ranch, CA, United States
Recall initiated2021-06-01
Classified2021-06-17
Reported by FDA2021-06-23
Terminated2023-06-22
DistributedNationwide (US)
Quantity10,469 items
Reason classespotency
Lot numbersMX180425
Model numbersMJS-101

Product

OSSUR MIAMI J SELECT COLLAR - Product Usage: is intended to provide gross immobilization to the cervical spine. REF: MJS-101; MJSR-101

Reason for recall

Due to insufficient column strength of the racks for all load conditions, racks associated with height adjustment functionality have been observed deforming, which may lead to increased range of motion in the cervical spine.