OSSUR MIAMI J SELECT COLLAR — Class II medical device recall Z-1896-2021
Terminated recall by Ossur Americas, reported 2021-06-23, distributed nationwide. Due to insufficient column strength of the racks for all load conditions, racks associated with height adjustm
| Recall number | Z-1896-2021 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Ossur Americas, Foothill Ranch, CA, United States |
| Recall initiated | 2021-06-01 |
| Classified | 2021-06-17 |
| Reported by FDA | 2021-06-23 |
| Terminated | 2023-06-22 |
| Distributed | Nationwide (US) |
| Quantity | 10,469 items |
| Reason classes | potency |
| Lot numbers | MX180425 |
| Model numbers | MJS-101 |
Product
OSSUR MIAMI J SELECT COLLAR - Product Usage: is intended to provide gross immobilization to the cervical spine. REF: MJS-101; MJSR-101
Reason for recall
Due to insufficient column strength of the racks for all load conditions, racks associated with height adjustment functionality have been observed deforming, which may lead to increased range of motion in the cervical spine.