Philips Allura Xper FD10 system with an ADN7NT patient table — Class II medical device recall Z-1896-2024
Ongoing recall by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V., reported 2024-06-05. When Software upgrade 8.1.100 is installed on an Allura Xper with an AD7NT (non tilt) table, the pivot brake d
| Recall number | Z-1896-2024 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Ongoing |
| Voluntary or mandated | Voluntary |
| Recalling firm | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V., Best, Netherlands |
| Recall initiated | 2024-05-16 |
| Classified | 2024-05-24 |
| Reported by FDA | 2024-06-05 |
| Countries outside the US | AE, ES, IN |
| Quantity | 3 units |
| Reason classes | software |
| UPC / GTIN / UDI-DI | 00884838059030 |
| Serial numbers | 526, 774, 909 |
| Model numbers | 722010 |
Product
Philips Allura Xper FD10 system with an ADN7NT patient table. Model Number: 722010, Software Version Number 8.1.100.
Reason for recall
When Software upgrade 8.1.100 is installed on an Allura Xper with an AD7NT (non tilt) table, the pivot brake does not engage. This issue may cause unexpected movements of the table.