Data Foundry

Medical device recalls 2012

PTS Panels Creatinine test strips — Class II medical device recall Z-1903-2012

Terminated recall by Polymer Technology Systems, Inc., reported 2012-07-04, distributed nationwide. During a routine product investigation, a creatinine sample with a value less than 1.0 mg/dl was tested on a v

Recall numberZ-1903-2012
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmPolymer Technology Systems, Inc., Indianapolis, IN, United States
Recall initiated2012-05-15
Classified2012-06-28
Reported by FDA2012-07-04
Terminated2012-07-20
DistributedNationwide (US)
Countries outside the USCN, FR, IE, PL
Quantity1,650 vials
Reason classespotency
Lot numbersF103
Model numbers1720

Product

PTS Panels Creatinine test strips. PTS Panels Creatinine Test Strips are intended to measure creatinine in whole blood. Creatinine measurements are used in the diagnosis and treatment of renal (kidney) diseases and in the monitoring of renal dialysis. This system is intended for the professional use for accurately and reproducibly measuring creatinine in human whole blood obtained from a finger stick sample

Reason for recall

During a routine product investigation, a creatinine sample with a value less than 1.0 mg/dl was tested on a version 2.60 CardioCek PA meter and the meter provided a result of 0.08mg/dl, which is not the reportable measuring range for the CardioChek PA meter for the creatinine assay.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1903-2012. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.