Data Foundry

Medical device recalls 2025

Broselow Pediatric Emergency Rainbow Tape — Class I medical device recall Z-1904-2025

Ongoing recall by SunMed Holdings, LLC, reported 2025-07-02, distributed nationwide. The impacted tape was manufactured with incorrect information on the tape. Incorrect values are printed in the

Recall numberZ-1904-2025
Product typeMedical device
ClassificationClass I
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmSunMed Holdings, LLC, Grand Rapids, MI, United States
Recall initiated2025-05-15
Classified2025-06-20
Reported by FDA2025-07-02
DistributedNationwide (US)
Countries outside the USCA
Quantity9,900 eaches
UPC / GTIN / UDI-DI10889483588970, 30889483588974
Lot numbers0004316661
Model numbers7700REA

Product

Broselow Pediatric Emergency Rainbow Tape (distribution by AirLife); REF 7700REA AirLife brand, 2025 Edition, and 36-23446 Rev 2 Print Version The Broselow Tape is a length- and weight-based reference tool specifically designed for pediatric emergency care.

Reason for recall

The impacted tape was manufactured with incorrect information on the tape. Incorrect values are printed in the Red zone, Orange zone, and Grey zone. Using the incorrect values in the Red zone could lead to shocking a patient with an excessive dose of joules, causing significant harm including burns, heart damage, and cardiac arrest. Incorrect values in the Orange zone could lead to overdosing the patient with sodium bicarbonate and cause metabolic alkalosis, electrolyte imbalances, tissue damage, and worsen respiratory status. Incorrect information in the Grey may lead to underdosing the patient with sodium bicarbonate and may cause reduced myocardial contractility, decreased response to vasopressors, and increased risk of dysrhythmia.