TDX SP Power Wheelchair — Class II medical device recall Z-1905-2012

Terminated recall by Invacare Corporation, reported 2012-07-11, distributed nationwide. Invacare Corporation decided to recall the product because of a potential risk of fire and serious injury. The

Recall numberZ-1905-2012
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmInvacare Corporation, Elyria, OH, United States
Recall initiated2012-04-06
Classified2012-06-29
Reported by FDA2012-07-11
Terminated2014-05-20
DistributedNationwide (US)
Countries outside the USAU, DE, NZ

Product

TDX SP Power Wheelchair, TDX SR Power Wheelchair, TDX SI Power Wheelchair. Model Number(s): TDXSI, TDXSI-2, TDXSI-2-S, TDXSI-CG, TDXSI-HD, TDXSI-HD-S, TDXSIV, TDXSIV-2, TDXSIV-2-S, TDXSIV-HD, TDXSIV-HD-S, TDXSP, TDXSPBASE, TDXSP-CG, TDXSPEURO, TDXSP-MCG, TDXSP-MCG-GT, TDXSR, TDXSR-CG, TDXSR-CG-HD, TDXSR-MCG, FDX, FDX-CG, FDXEURO-14, FDX-MCG. Mrf. By: Invacare Corporation, 1200 Taylor Street, Elyria, OH 44035. Intended to provide mobility to a person restricted to a sitting position.

Reason for recall

Invacare Corporation decided to recall the product because of a potential risk of fire and serious injury. The post fuse assemblies could potentially have a short between the positive ring terminal and the terminal battery bracket which would create an alternate current path around the fuse.