TDX SP Power Wheelchair — Class II medical device recall Z-1905-2012
Terminated recall by Invacare Corporation, reported 2012-07-11, distributed nationwide. Invacare Corporation decided to recall the product because of a potential risk of fire and serious injury. The
| Recall number | Z-1905-2012 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Invacare Corporation, Elyria, OH, United States |
| Recall initiated | 2012-04-06 |
| Classified | 2012-06-29 |
| Reported by FDA | 2012-07-11 |
| Terminated | 2014-05-20 |
| Distributed | Nationwide (US) |
| Countries outside the US | AU, DE, NZ |
Product
TDX SP Power Wheelchair, TDX SR Power Wheelchair, TDX SI Power Wheelchair. Model Number(s): TDXSI, TDXSI-2, TDXSI-2-S, TDXSI-CG, TDXSI-HD, TDXSI-HD-S, TDXSIV, TDXSIV-2, TDXSIV-2-S, TDXSIV-HD, TDXSIV-HD-S, TDXSP, TDXSPBASE, TDXSP-CG, TDXSPEURO, TDXSP-MCG, TDXSP-MCG-GT, TDXSR, TDXSR-CG, TDXSR-CG-HD, TDXSR-MCG, FDX, FDX-CG, FDXEURO-14, FDX-MCG. Mrf. By: Invacare Corporation, 1200 Taylor Street, Elyria, OH 44035. Intended to provide mobility to a person restricted to a sitting position.
Reason for recall
Invacare Corporation decided to recall the product because of a potential risk of fire and serious injury. The post fuse assemblies could potentially have a short between the positive ring terminal and the terminal battery bracket which would create an alternate current path around the fuse.