Data Foundry

Medical device recalls 2025

Dexcom G7 iOS Application — Class II medical device recall Z-1906-2025

Ongoing recall by Dexcom, Inc., reported 2025-06-11, distributed nationwide. A bug in the 2.7.0 version of the G7 iOS App can cause the app to enter a state where it stops updating the es

Recall numberZ-1906-2025
Product typeMedical device
ClassificationClass II
StatusOngoing
Voluntary or mandatedVoluntary
Recalling firmDexcom, Inc., San Diego, CA, United States
Recall initiated2025-05-05
Classified2025-06-04
Reported by FDA2025-06-11
DistributedNationwide (US)
Countries outside the USAE, AT, AU, BE, BH, CA, CH, CZ, DE, DK, ES, FI, FR, GB, HK, IE, IL, IS, IT, JO, KR, KW, LB, LT, MT, NL, NO, NZ, OM, PL, PT, QA, SA, SE, SI, SK, TR, ZA
Quantity61,355 users
UPC / GTIN / UDI-DI00386270001863
Model numbersSW12300

Product

Dexcom G7 iOS Application, Ref: SW12300 used in conjunction with the Dexcom G7 Continuous Glucose Monitoring (CGM) system

Reason for recall

A bug in the 2.7.0 version of the G7 iOS App can cause the app to enter a state where it stops updating the estimated glucose value (EGV) and continues to show this stale EGV. If the user is unaware that the EGVs are not being updated and that glucose alerts will not trigger, there is potential for the missed detection of a hyperglycemic or hypoglycemic event or a treatment decision made based off incorrect data. There is no impact to transmitter communication with concurrently connected displays. A concurrently connected receiver, Apple Watch, and/or insulin pump will continue to receive EGVs directly from the transmitter when the app is in the problematic state. There is no impact to Automated Insulin Dosing (AID) therapy as the AID system continues to receive EGVs directly from the transmitter.