Data Foundry

Medical device recalls 2012

ULTIMA*LX STM — Class II medical device recall Z-1910-2012

Terminated recall by DePuy Orthopaedics, Inc., reported 2012-07-11, distributed nationwide. DePuy Orthopaedics, Inc. is issuing a Field Safety Notice for the LUSTER Cemented Polished Femoral Hip Stems a

Recall numberZ-1910-2012
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmDePuy Orthopaedics, Inc., Warsaw, IN, United States
Recall initiated2012-05-16
Classified2012-06-29
Reported by FDA2012-07-11
Terminated2013-03-05
DistributedNationwide (US)
Quantity431 ultima
Lot numbers438752, 475234, 633707, XBK74

Product

ULTIMA*LX STM,POL 10/12TAP/SZ5 Packaging: The product was packaged within inner and outer PETG blister trays and sealed with coated Tyvek lids. Foam protectors were in direct contact with the stem to help immobilize the product within the package. The sterile barrier system was then packed within a carton and shrink wrapped. Intended Use: The ULTIMA LX Cemented Polished Femoral Hip Stems product are forged CoCr polished cemented stems which are used as part of a total hip arthroplasty. These stems were indicated for use with all DePuy bearings and acetabular cups.

Reason for recall

DePuy Orthopaedics, Inc. is issuing a Field Safety Notice for the LUSTER Cemented Polished Femoral Hip Stems and the ULTIMA LX Cemented Polished Femoral Hip Stems that were implanted with any metal-on-metal (MoM) articulation (including a head of CoCr and an acetabular cup where the cup is of an all metal construction (monoblock or modular)).The LUSTER Cemented Polished Femoral Hip Stems and the U