Data Foundry

Medical device recalls 2016

VITROS 3600 Immunodiagnostic System — Class II medical device recall Z-1913-2016

Terminated recall by Ortho-Clinical Diagnostics, reported 2016-06-15, distributed nationwide. Ortho Clinical Diagnostics confirmed a software timing anomaly in which the VITROS 3600 and 5600 Systems may u

Recall numberZ-1913-2016
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmOrtho-Clinical Diagnostics, Rochester, NY, United States
Recall initiated2016-04-06
Classified2016-06-06
Reported by FDA2016-06-15
Terminated2018-06-01
DistributedNationwide (US)
Countries outside the USAU, BR, CA, CL, CN, CO, DE, ES, FR, GB, IN, IT, JP, MX, PA, SG, VE
Reason classessoftware
UPC / GTIN / UDI-DI10758750002979
Serial numbersJ36000012, J36000933.UNIQUE
Model numbers6802783

Product

VITROS 3600 Immunodiagnostic System, Catalog Number 6802783,for use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Immunodiagnostic Products Reagents.

Reason for recall

Ortho Clinical Diagnostics confirmed a software timing anomaly in which the VITROS 3600 and 5600 Systems may unexpectedly aspirate from or dispense into an unintended sample that may lead to erroneous or contaminated results.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1913-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.