Cholesterol CHOD-PAP — Class II medical device recall Z-1920-2015

Terminated recall by Roche Diagnostics Operations, Inc., reported 2015-07-08, distributed nationwide. Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trin

Recall numberZ-1920-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmRoche Diagnostics Operations, Inc., Indianapolis, IN, United States
Recall initiated2015-05-07
Classified2015-06-27
Reported by FDA2015-07-08
Terminated2017-03-24
DistributedNationwide (US)
Quantity6,255 analytical
Model numbers11491458216

Product

Cholesterol CHOD-PAP. In vitro diagnostic reagent system intended for use on the Hitachi systems for The quantitative determination of total cholesterol in serum and plasma.

Reason for recall

Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.

Machine access

This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1920-2015. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.

Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.