ROHO HIGH PROFILE Single Compartment Cushion with Sensor Ready… — Class II medical device recall Z-1921-2017

Terminated recall by Roho Inc., reported 2017-05-10, distributed nationwide. When the Smart Check tool is used to perform a Smart Setup with one of the affected cushions, it may adversely

Recall numberZ-1921-2017
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmRoho Inc., Belleville, IL, United States
Recall initiated2017-03-20
Classified2017-05-02
Reported by FDA2017-05-10
Terminated2019-05-10
DistributedNationwide (US)
Countries outside the USAE, AT, AU, CA, CH, CZ, DE, ES, FI, FR, GB, HK, IE, IS, IT, JP, KR, LU, NZ, SE, SG, ZA
Quantity5,646 units
Serial numbersA2920163
Model numbers1R1010H-CA-SC, 1R1010H-CA-SCDM, 1R1010H-CA-SR, 1R1010H-CF-SCDM, 1R1010H-CF-SR, 1R1010H-CH-SR, 1R1010M-CA-SC, 1R1010M-CA-SCDM, 1R1010M-CA-SR, 1R1010M-CF-SCDM, 1R1010M-CF-SR, 1R1010M-CH-SR, 1R1011H-CA-SC, 1R1011H-CA-SR, 1R1011H-CF-SR, 1R1011H-CH-SC, 1R1011H-CH-SR, 1R1011M-CA-SC, 1R1011M-CA-SR, 1R1011M-CF-SR, 1R1012H-CA-SR, 1R1012H-CF-SR, 1R1012H-CH-SR, 1R108H-CA-SR, 1R108H-CF-SR and 97 more

Product

ROHO HIGH PROFILE Single Compartment Cushion with Sensor Ready Technology and ROHO MID PROFILE Single Compartment Cushion with Sensor Ready Technology.

Reason for recall

When the Smart Check tool is used to perform a Smart Setup with one of the affected cushions, it may adversely affect the setup process resulting in an improper cushion setup. Sitting on an over inflated or under inflated Sensor Ready Cushion may reduce or eliminate the benefits provided by the cushion, increasing the risk to skin or to other soft tissue.