Siemens Dimension LOCI TNI Troponin I Calibrator RC 621 — Class II medical device recall Z-1926-2014
Terminated recall by Siemens Healthcare Diagnostics, Inc., reported 2014-07-02, distributed nationwide. Siemens Dimension LOCI TNI Troponin I calibrator lot 3DD007 may demonstrate an average upward shift of 24% in
| Recall number | Z-1926-2014 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Siemens Healthcare Diagnostics, Inc., Newark, DE, United States |
| Recall initiated | 2013-09-17 |
| Classified | 2014-06-26 |
| Reported by FDA | 2014-07-02 |
| Terminated | 2014-08-27 |
| Distributed | Nationwide (US) |
| Quantity | 1,625 |
| Lot numbers | 3DD007 |
Product
Siemens Dimension LOCI TNI Troponin I Calibrator RC 621. The LOCI TNI Troponin I Calibrator is an in vitro diagnostic product for the calibration of the cardiac troponin I (TNI) method for the Dimension clinical EXL integrated chemistry system with LOCI module.
Reason for recall
Siemens Dimension LOCI TNI Troponin I calibrator lot 3DD007 may demonstrate an average upward shift of 24% in QC and patient results following calibration.