Medtronic RestoreSensor — Class II medical device recall Z-1926-2016
Terminated recall by Medtronic Neuromodulation, reported 2016-06-15. Loss of therapy during recharging of device. Reportedly, a charging session was terminated prior to obtaining
| Recall number | Z-1926-2016 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Medtronic Neuromodulation, Minneapolis, MN, United States |
| Recall initiated | 2016-05-03 |
| Classified | 2016-06-07 |
| Reported by FDA | 2016-06-15 |
| Terminated | 2017-08-04 |
| Quantity | 60,941 |
| Model numbers | 37714, 97714 |
Product
Medtronic RestoreSensor, Model 37714; and, RestoreSensor SureScan, (MRI), Model 97714, Implantable Neurostimulators. Used to treat chronic, intractable pain of the trunk and/or limbs-including unilateral or bilateral pain.
Reason for recall
Loss of therapy during recharging of device. Reportedly, a charging session was terminated prior to obtaining recharge threshold voltage of 3.615, which triggered rapid battery depletion state. As a result, the device depleted to 1.925 volts in one to two days rather than the typical 30 days. Insufficient coupling between recharger and device during recharge session was found to be key factor.
Machine access
This record, with every extracted code, the verbatim FDA record and provenance, is available from the recall API: GET https://api.data.aroqon.com/v1/recalls/Z-1926-2016. Look up any UPC, UDI, NDC or lot with /v1/recalls/lookup?code=…. API docs · free and paid keys.
Source: U.S. Food and Drug Administration enforcement report via openFDA (CC0). Structured by Data Foundry (recall-structuring@2); last changed 2026-09-26. Not affiliated with or endorsed by FDA. Not medical or legal advice.