Triglycerides GPO — Class II medical device recall Z-1929-2015
Terminated recall by Roche Diagnostics Operations, Inc., reported 2015-07-08, distributed nationwide. Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trin
| Recall number | Z-1929-2015 |
|---|---|
| Product type | Medical device |
| Classification | Class II |
| Status | Terminated |
| Voluntary or mandated | Voluntary |
| Recalling firm | Roche Diagnostics Operations, Inc., Indianapolis, IN, United States |
| Recall initiated | 2015-05-07 |
| Classified | 2015-06-27 |
| Reported by FDA | 2015-07-08 |
| Terminated | 2017-03-24 |
| Distributed | Nationwide (US) |
| Quantity | 6,255 |
| Model numbers | 04657594190 |
Product
Triglycerides GPO. In vitro test for the quantitative determination of triglycerides in human serum and plasma on the COBAS Integra 800 and 400 plus systems, cobas c111, Roche/Hitachi MODULAR systems, and cobas c systems.
Reason for recall
Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.