Triglycerides/Glycerol Blanked — Class II medical device recall Z-1930-2015

Terminated recall by Roche Diagnostics Operations, Inc., reported 2015-07-08, distributed nationwide. Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trin

Recall numberZ-1930-2015
Product typeMedical device
ClassificationClass II
StatusTerminated
Voluntary or mandatedVoluntary
Recalling firmRoche Diagnostics Operations, Inc., Indianapolis, IN, United States
Recall initiated2015-05-07
Classified2015-06-27
Reported by FDA2015-07-08
Terminated2017-03-24
DistributedNationwide (US)
Quantity6,255
Model numbers11877771216

Product

Triglycerides/Glycerol Blanked. In vitro test for the quantitative determination of triglycerides in human serum and plasma or Roche/Hitachi MODULAR systems, and cobas c systems.

Reason for recall

Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.